![]() ![]() ![]() When the probe is disconnected and a Fluke model 87 DMM (specified to have 1 volt accuracy on the 0 - 1000 VDC range) is used to measure the meter HV the result is 695 Volts, not the 900 volts mentioned in the sales literature for this model. The PGM pancake probe housing has a 55 mm window opening which has a window screen covering and the GM tube is rated as having an effective opening of 44.5 mm. web page only the 7311 has the correct voltage rating and physical configuration that will fit the PGM housing and is rated for 900 volt operation. Of all the pancake probes shown on the LND, Inc. ![]()